Temperature quantification using the proton frequency shift technique: In vitro and in vivo validation in an open 0.5 tesla interventional MR scanner during RF ablation
René M. Botnar,Paul Steiner,Benjamin Dubno,Peter Erhart,Gustav K. von Schulthess,Jörg F. Debatin +5 more
53
TL;DR: The proton frequency shift (PFS) technique was used to synthesize thermal maps after RF ablation in an open 0.5 T MR system under ex vivo and in vivo conditions using the PFS technique, an accurate and fast imaging method for temperature quantification.
read more
Abstract: Open magnetic resonance (MR) scanners allow MR-guided targeting of tumors, as well as temperature monitoring of radio frequency (RF) ablation. The proton frequency shift (PFS) technique, an accurate and fast imaging method for temperature quantification, was used to synthesize thermal maps after RF ablation in an open 0.5 T MR system under ex vivo and in vivo conditions. Calibration experiments with 1.5% agarose gel yielded a chemical shift factor of 0.011 +/- 0.001 ppm/ degrees C (r2 = 0.96). Three gradient echo (GRE) pulse sequences were tested for thermal mapping by comparison with fiberoptic thermometer (Luxtron Model 760) readings. Temperature uncertainty decreased from high to low bandwidths (BW): +/-5.9 degrees C at BW = 15.6 kHz, +/-1.4 degrees C at BW = 3.9 kHz, and +/-0.8 degrees C at BW = 2.5 kHz. In vitro experiments (N = 9) in the paraspinal muscle yielded a chemical shift factor of 0.008 +/- 0.001 ppm/ degrees C. Temperature uncertainty was determined as +/-2.7 degrees C (BW = 3.9 kHz, TE = 19.3 msec). The same experiments carried out in the paraspinal muscle (N = 9) of a fully anesthetized pig resulted in a temperature uncertainty of +/-4.3 degrees C (BW = 3.9 kHz, TE = 19.3 msec), which is higher than it is in vitro conditions (P < 0.15). Quantitative temperature monitoring of RF ablation is feasible in a 0.5 T open-configured MR scanner under ex vivo and in vivo conditions using the PFS technique.
read more
Chat with Paper
AI Agents for this Paper
Find similar papers on Google Scholar, PubMed and Arxiv
Write a critical review of this paper
Analyze citations of this paper to find unaddressed research gaps
Citations
Quantitative MRI-based temperature mapping based on the proton resonant frequency shift: Review of validation studies
TL;DR: MRI-based temperature imaging that exploits the temperature-sensitive water proton resonant frequency shift is the only available method for reliable quantification of temperature changes in vivo and clinical studies that have shown this method feasible in human treatments are overviewed.
225
Magnetic resonance temperature imaging
TL;DR: Real-time image processing and visualization techniques, together with accelerated MRI acquisition techniques, are described because of their potential for therapy guidance.
Magnetic resonance guidance for radiofrequency ablation of liver tumors.
TL;DR: The role of MR guidance for RF ablation of liver tumors is summarized, which offers the possibility for a safe and effective therapy option in the treatment of primary and secondary hepatic malignancies.
115
Noninvasive Magnetic Resonance Thermography of Recurrent Rectal Carcinoma in a 1.5 Tesla Hybrid System
Johanna Gellermann,Waldemar Wlodarczyk,Bert Hildebrandt,Hildegard Ganter,Anett Nicolau,Beate Rau,Wolfgang Tilly,H. Fähling,Jacek Nadobny,Roland Felix,Peter Wust +10 more
TL;DR: In 15 patients with preirradiated pelvic rectal recurrences, the mean T(MR) of the tumor might be useful as a variable to evaluate the quality and effectivity of heat treatments, and consequently as optimization variable.
Accuracy of real-time MR temperature mapping in the brain: a comparison of fast sequences.
TL;DR: The ss EPI sequence was the fastest and the most precise sequence for MR thermometry, with significantly higher accuracy compared to GRE.
68
References
•Book
Principles of Nuclear Magnetic Resonance Microscopy
Paul T. Callaghan
- 01 Jan 1991
TL;DR: In this paper, the authors introduce the principles of imaging for nuclear magnetic resonance and describe the influence of magnetic field gradients on the magnetic field. But they do not discuss the application of magnetic resonance in biology and minerals science.
Principles of Nuclear Magnetic Resonance Microscopy
TL;DR: In this paper, the authors group together two apparently very different applications of magnetic field gradients, showing the common physical principles which determine the imaging of static density distributions via k-space and dynamic displacements via ''q-space'' acquisition.
2.4K
Small Hepatocellular Carcinoma: Treatment with Radio-frequency Ablation versus Ethanol Injection
Tito Livraghi,S. Nahum Goldberg,Sergio Lazzaroni,Franca Meloni,Luigi Solbiati,G. Scott Gazelle +5 more
TL;DR: RF ablation results in a higher rate of complete necrosis and requires fewer treatment sessions than percutaneous ethanol injection, however, the complication rate is higher with RF ablation than with percutaneously ethanol injection.
1.2K
A precise and fast temperature mapping using water proton chemical shift.
Yasutoshi Ishihara,Calderon A,Hidehiro Watanabe,Kazuya Okamoto,Yoshinori Suzuki,Kagayaki Kuroda +5 more
TL;DR: A new temperature measurement procedure using phase mapping was developed that makes use of the temperature dependence of the water proton chemical shift and highly accurate and fast measurements were obtained during phantom and in vivo experiments.
1K
Proton Resonance Shift of Water in the Gas and Liquid States
TL;DR: In this paper, a consistent treatment of the chemical shift, thermal and dielectric data for water can be given based on a two-state model involving an equilibrium between a hydrogen-bonded ''icelike'' fraction and a ''monomer'' fraction whose interaction with the lattice arises entirely from London dispersion forces.
689