Modern Pacemaker and Implantable Cardioverter/Defibrillator Systems Can Be Magnetic Resonance Imaging Safe In Vitro and In Vivo Assessment of Safety and Function at 1.5 T
Ariel Roguin,Menekhem M. Zviman,Glenn Meininger,E. Rene Rodrigues,Timm Dickfeld,David A. Bluemke,Albert C. Lardo,Ronald D. Berger,Hugh Calkins,Henry R. Halperin +9 more
393
TL;DR: These data suggest that certain modern pacemaker and ICD systems may indeed be MRI safe, and may have major clinical implications for current imaging practices.
read more
Abstract: Background— MRI has unparalleled soft-tissue imaging capabilities. The presence of devices such as pacemakers and implantable cardioverter/defibrillators (ICDs), however, is historically considered a contraindication to MRI. These devices are now smaller, with less magnetic material and improved electromagnetic interference protection. Our aim was to determine whether these modern systems can be used in an MR environment. Methods and Results— We tested in vitro and in vivo lead heating, device function, force acting on the device, and image distortion at 1.5 T. Clinical MR protocols and in vivo measurements yielded temperature changes <0.5°C. Older (manufactured before 2000) ICDs were damaged by the MR scans. Newer ICD systems and most pacemakers, however, were not. The maximal force acting on newer devices was <100 g. Modern (manufactured after 2000) ICD systems were implanted in dogs (n=18), and after 4 weeks, 3- to 4-hour MR scans were performed (n=15). No device dysfunction occurred. The images were o...
read more
Chat with Paper
AI Agents for this Paper
Find similar papers on Google Scholar, PubMed and Arxiv
Write a critical review of this paper
Analyze citations of this paper to find unaddressed research gaps
Citations
Practice advisory for the perioperative management of patients with cardiac rhythm management devices: Pacemakers and implantable cardioverter- defibrillators - A report by the American Society of Anesthesiologists Task Force on perioperative management of patients with cardiac rhythm management devices
James R. Zaidan,John L. Atlee,Peter Belott,Kurt S. Briesacher,Richard T. Connis,John D. Gallagher,David Haves,Jane E. Hershey,Neal Kay,David G. Nickinovich,Marc A. Rozner,Mark F. Trankina +11 more
TL;DR: Practice advisories summarize the state of the literature and report opinions derived from a synthesis of task force members, expert consultants, open forums, and public commentary.
149
Advanced ICD troubleshooting: Part II.
TL;DR: Failure to Deliver Therapy or Delay of Therapy Absent or delayed therapy may be caused by programmed values (including human error), ICD system performance, or a combination of the two.
137
A Technique to Evaluate MRI-Induced Electric Fields at the Ends of Practical Implanted Lead
TL;DR: In this article, the reciprocity theorem in conjunction with the Huygens principle is used to decouple the micro-scale metallic lead simulation/measurement from the macro-level phantom human simulations within the MRI scanners.
Magnetic resonance imaging at 1.5-T in patients with implantable cardioverter-defibrillators.
Claas P. Naehle,K. Strach,Daniel Thomas,Carsten Meyer,Markus Linhart,Sascha Bitaraf,Harold Litt,Jörg O. Schwab,Hans H. Schild,Torsten Sommer +9 more
TL;DR: MRI of non-pacemaker-dependent ICD patients can be performed with an acceptable risk/benefit ratio under controlled conditions by taking both MRI- and pacemaker-related precautions.
127
References
The use of contrast-enhanced magnetic resonance imaging to identify reversible myocardial dysfunction.
Raymond J. Kim,Edwin Wu,Allen Rafael,Enn-Ling Chen,Michele Parker,Orlando P. Simonetti,Francis J. Klocke,Robert O. Bonow,Robert M. Judd +8 more
TL;DR: Reversible myocardial dysfunction can be identified by contrast-enhanced MRI before coronary revascularization and is strongly related to the degree of improvement in the global mean wall-motion score and the ejection fraction after Revascularization.
3.3K
Magnetic Resonance Imaging Determination of Cardiac Prognosis
W. Gregory Hundley,Timothy M. Morgan,Christina M. Neagle,Craig A. Hamilton,Pairoj Rerkpattanapipat,Kerry M. Link +5 more
TL;DR: In patients with poor echocardiograms, the results of cardiac MRI stress tests can be used to forecast myocardial infarction or cardiac death, independent of the presence of risk factors for coronary arteriosclerosis.
308
Electromagnetic interference in patients with implanted pacemakers or cardioverter-defibrillators
TL;DR: It is accepted practice to advise patients that while airport screening devices may detect the pacemaker or ICD metal case the device will not be adversely affected, and patients should carry their device identification card for the purpose of obtaining security clearance.
136
Magnetic resonance imaging safety: implications for cardiovascular patients.
Salmaan Ahmed,Frank G. Shellock +1 more
TL;DR: The bioeffects of MR exposures are discussed and an overview of safety considerations are provided, with an emphasis on information pertinent to cardiovascular patients.
83